METAMAX® 3B

Portable. Wearable. Wireless.

Portable CPET system: flexible to use in the lab and in the field.

Mobile CPET in diagnostic quality: the METAMAX® 3B measures Breath-by-Breath wherever you test, in the lab or out in the field.

METAMAX® 3B portable breath gas analyzer

The complete measurement system weighs 580 grams, is worn on the body, and runs wirelessly on battery power. The METAMAX® 3B transmits every measurement telemetrically over more than 1,000 meters or stores it autonomously on board. Its range of applications extends from cardiac rehabilitation to the ISS.

Highlights

Mobile CPET without compromise.

  • MMX in stationary use

    Every breath analyzed

    The system captures oxygen uptake, carbon dioxide output, and breathing volume for every single breath in real time. This high temporal resolution is the basis for assessing V̇O₂ kinetics and ventilatory thresholds, in the lab or out in the field.

  • MMX antenna

    Range of more than 1,000 meters

    Bidirectional telemetry transmits the data to the PC in real time via Long Range Bluetooth® with a range of more than 1,000 m, even without line of sight and without cables.

  • MMX 3B Buttons

    On-device operation

    PC or standalone: the system is operated directly on the device or via MetaSoft® Studio on a laptop.

  • MMX 3B R2 Indoor Cyclist

    One Device, Twice the Benefit

    The optional Stationary Kit turns the mobile analyzer into a laboratory workstation with an external power supply and an extended sample line. One device for two fields of application.

  • MMX paragliding

    Featherlight in use

    At 580 g, worn in a chest or back carrying system, the device weight places practically no limit on the test subject during the test. The fit is free of pressure points.

  • MMX Elevation

    Ready for extreme conditions

    Whether in snow and ice at down to -10 °C or in the desert at temperatures of up to 40 °C, the METAMAX® 3B is insensitive to extreme outdoor temperatures.

Where standard diagnostics can't go.

The METAMAX® 3B measures under conditions laboratory systems were never built for. In the water, under respiratory protection, under specialized equipment.

Swimmer doing front crawl in a swim flume, connected to the breath gas measurement of the METAMAX® 3B via the Swim Option
01SwimmingWith the swimming option, the METAMAX® 3B also measures in the water: in the swim flume, in the pool, and in open water.
Firefighters wearing breathing apparatus and full gear during an operational drill, exercise testing with the METAMAX® 3B
02Fire services &
occupational medicine
Exercise testing under real conditions: the METAMAX® 3B captures the combined influence of heat, stress, respiratory protection, and heavy equipment directly during operational training.
Pilot in a flight suit with breathing mask, connected to the METAMAX® 3B for aeromedical exercise testing
03Aviation medicineAviation medicine in action: the METAMAX® 3B makes the physical strain inside a flight suit measurable.

MetaSoft® Studio

Reliably connected

An examination runs only as smoothly as the software behind it. MetaSoft® Studio controls the entire test workflow, from load protocol to connected devices to test report.

Every measurement is preserved as a raw data set and remains traceable. You focus on the examination while the technology works in the background.

Whether in the lab, in the practice, or in the field: it is the same software with the same workflow, and all results land in the same evaluation.

MetaSoft® Studio, test view with breathing gas curves, workload, and live readings

Clicking loads content from YouTube · Privacy notice

Missions

Mission logo "Metabolic Space on ISS", CORTEX & TU Dresden, 2018

On a space mission

In the summer of 2018, German astronaut Dr. Alexander Gerst set out on a research mission aboard the ISS. On board for the first time was a mobile CPET device, our METAMAX® 3B. During his stay on the space station, Alexander Gerst used it several times to test his health status and fitness.

Studies with the METAMAX® 3B

The 5 latest publications from research and practice.

View all studies
Specifications

Measurement

Measuring method
Breath-by-Breath
Flow (turbine)
Reusable: ≤ 20 L/s, ± 2%Disposable: ≤ 14 L/s, ± 3% or ± 50 mL (whichever is greater)

Calibration & maintenance

Gas sensors
monthly, two-point
Reusable turbine
daily or before each test
Disposable turbine
factory-calibrated
Air pressure
every six months
Sensor warm-up time
30 min
O₂ sensor
Replacement recommended after 6 months

Interfaces

PC
Long Range Bluetooth® receiver (USB)
Radio standard
Bluetooth® 2.1 + EDR, 2.4 GHz
Telemetry range
max. over 1,000 m
Heart rate
Bluetooth® Smart HR strap
SpO₂
Nonin Bluetooth® 3150 Wrist pulse oximeter
ECG
14 ECG systems, includingGE CardioSoft, Schiller, Spacelabs Healthcare

Environment

Operating temperature
-10 °C to +40 °C
Humidity
≤ 95% relative humidity, non-condensing
Ambient pressure
500–1,050 mbar

Device & operation

Dimensions
2 housing shells, each 120 × 110 × 45 mm
Weight (without battery)
580 g
Battery life
approx. 5.5 h per battery
Operating modes
PC operationstandalone without PC

Approval

Class
Class IIa medical device according to 93/42/EEC
Marking
CE marking

Subject to technical changes.

FAQ
How far does the wireless transmission really reach?

The bidirectional Long Range Bluetooth® receiver is designed for ranges of more than 1,000 meters. In practice, the effective range depends on line of sight and surroundings. The METAMAX® 3B can also store measurement data in its internal data logger, independently of the PC. After the test, the data is transferred to the PC via Bluetooth® and evaluated in MetaSoft® Studio.

How often does the system need to be calibrated?

The gas sensors are adjusted once a month by two-point calibration with ambient air and calibration gas, and the gas-on-demand valve doses the calibration gas automatically. You calibrate the reusable turbine daily or before each test, while disposable turbines are factory-calibrated. Air pressure calibration is only scheduled every six months. The device warm-up time is 30 minutes.

Can the METAMAX® 3B also be used in stationary mode?

Yes. With the optional Stationary Kit, the mobile analyzer becomes a laboratory workstation with an external power supply and an extended sample line. One device, two fields of application.

Can I connect the METAMAX® 3B to an ECG?

Yes. Via MetaSoft® Studio, the METAMAX® 3B connects to 14 ECG systems, including GE CardioSoft, Schiller, and Spacelabs Healthcare. You can find the complete list on the METALYZER® 3B page.

What training is required?

Every delivery includes installation and introductory training. We offer in-depth user training on site, in hybrid form, or in Leipzig. Training partner: ZfS Münster.